Tandem Mobi Insulin Pump Recall: FDA Flags 1,248 Serious Injuries. How to Check Your Pump and Stay Safe

Date:

If you or your child wears a Tandem Mobi insulin pump, take a few minutes today to look up its serial number. On November 25, 2026, the FDA labeled the Tandem Mobi recall its most serious type, after Tandem Diabetes Care reported 1,248 serious injuries linked to false blockage alarms that can pause insulin delivery. No deaths have been reported.

The recall covers 80,938 pumps, 77,559 of them in the United States, according to the FDA’s recall database. Most owners can keep using their pump under updated instructions. A smaller group, whose pumps have already shown frequent false alarms, is being told to get a replacement right away. Here is how to tell which group you’re in and what to do in either case.

Key Takeaways

  • What goes wrong: Some Tandem Mobi pumps sound a false occlusion (blockage) alarm and stop delivering insulin until the user responds, even when nothing is blocked.
  • How serious: The FDA classified it as the most serious recall type on Nov. 25. Tandem has reported 1,248 serious injuries and no deaths as of Oct. 28, 2026.
  • Two groups of pumps: About 4,600 serial numbers on the FDA list should be replaced now. All other Tandem Mobi pumps with Interoperable Technology can stay in use with new instructions.
  • Never ignore an alarm: Acknowledge it, check the cartridge, tubing and infusion site, fix any problem, then resume insulin.
  • Have a plan B: Keep backup insulin supplies ready and turn on push notifications in the Tandem Mobi app.
  • Who to call: Tandem Customer Technical Support at 1-877-801-6901.

What Is Wrong With the Tandem Mobi Insulin Pump?

An insulin pump watches for blockages in its tubing, because a kinked line or clogged site means insulin isn’t getting through. Tandem found that in certain Mobi pumps, extra friction inside the drive module can sap the motor’s power. Under some conditions the motor slows below the speed the pump uses to detect a blockage, so the pump decides there is an occlusion when there isn’t one.

That false alarm matters because of what happens next. Insulin delivery stays suspended until the user acknowledges the alarm and restarts therapy. If alarms keep happening and aren’t handled quickly, the FDA says the interruptions can drive blood sugar up, and in severe cases hyperglycemia may need medical treatment or a hospital stay. The FDA’s database lists the cause as a problem with process change control.

Tandem first notified customers by email on September 23, 2026. The FDA posted the action in its database on October 30 and published its formal safety notice, with the most serious classification, on November 25.

Which Tandem Mobi Pumps Are Included in the Recall?

Every Tandem Mobi Insulin Pump with Interoperable Technology is part of this recall, but the instructions differ depending on the serial number.

If your pump is…What Tandem and the FDA recommend
On the replacement list (about 4,600 serial numbers with frequent false alarms)Request a return and replacement immediately, using the link in Tandem’s letter or by calling 1-877-801-6901. Until the new pump arrives, follow the safety steps below and check your blood sugar regularly.
Not on the replacement listKeep using it while following the updated instructions. Call Tandem if false alarms become unusually frequent or keep disrupting therapy after troubleshooting.
Source: FDA safety notice and recall database entry Z-0316-2027 (Nov. 2026).

The serial number list is long. The quickest way to check is to open the FDA recall entry on a computer, press Ctrl+F (Command+F on a Mac) and type your seven-digit serial number. TechRepublic’s walkthrough notes that the Tandem Mobi app shows the serial number under Settings, then Pump, then Pump Info. If you can’t find it or aren’t sure, call Tandem and ask them to check for you.

What Should Tandem Mobi Users Do Right Now?

Whatever group your pump falls into, these steps come straight from the FDA’s notice and Tandem’s letters:

  1. Respond to every occlusion alarm. Acknowledge it, inspect the cartridge, tubing and infusion site for damage or blockage, correct anything you find and resume insulin delivery. Do not assume an alarm is false.
  2. Turn on push notifications in the Tandem Mobi app so your phone alerts you too. The pump itself uses tones or vibration, getting louder until you respond. Tandem has a support article on enabling app alerts.
  3. Check your blood sugar regularly, especially if you are still using a pump that is on the replacement list.
  4. Keep a backup method of insulin on hand, such as pens or syringes, as your care team prescribed during pump training.
  5. Call Tandem if alarms are unusually frequent, keep coming back after troubleshooting, or you need help clearing one.
  6. Fill out the acknowledgment form linked in Tandem’s letter so the company knows you received the notice.

Don’t stop using your pump on your own without a backup plan in place. Going without insulin is the very risk this recall is trying to prevent. If you’re unsure, call your endocrinologist or diabetes educator before you change anything.

Caregivers: check the pumps you help manage

Many children, teens and older adults rely on a parent, spouse or aide to handle alarms, especially overnight. Make sure whoever helps knows how to respond to an occlusion alarm, has the Tandem app with notifications on, and knows where the backup insulin is kept. Our guide to checking medicines and devices after a recall has more tips for caregivers.

What Are the Warning Signs of High Blood Sugar After a Pump Pause?

When a pump stops delivering insulin, blood sugar can climb within hours, and people with type 1 diabetes are most at risk of diabetic ketoacidosis (DKA). The CDC lists missed insulin and a clogged insulin pump among the most common causes of DKA. Early signs include strong thirst and urinating much more than usual.

Later symptoms can come on fast: fast, deep breathing, fruity-smelling breath, nausea or vomiting, stomach pain, a flushed face, headache and deep tiredness. The CDC advises checking blood sugar every four to six hours and testing for ketones when blood sugar is 250 mg/dL or higher. Call 911 or go to the emergency room if blood sugar stays at 300 mg/dL or above, your breath smells fruity, you’re vomiting and can’t keep fluids down, or you’re having trouble breathing.

Why Did the FDA Call This Recall Its Most Serious Type?

The FDA uses that label when continued use of a device without following the recommended steps could cause serious injury or death. It doesn’t mean every pump is failing. It means the consequences of a missed alarm can be severe for someone who depends on steady insulin, which is why the agency is pushing every Mobi user to act, not just people on the replacement list.

This isn’t Tandem’s first Mobi problem. A separate software issue that could affect insulin delivery led the company to send a fix to customers in October 2025, MedTech Dive reported. For other recent diabetes treatment news, see our report on the FDA’s approval of once-weekly insulin for children with type 2 diabetes.

How Do I Report a Problem With My Insulin Pump?

Report it to Tandem first at 1-877-801-6901 or techsupport@tandemdiabetes.com so the company can evaluate your pump. You or your clinician can also file a report with MedWatch, the FDA’s safety reporting program, or call 1-888-INFO-FDA. These reports help the FDA spot patterns, and they are how injury counts like the 1,248 in this recall come to light.

Traveling for the holidays? Pack extra infusion sets, cartridges and backup insulin in your carry-on, and keep your phone charged so app alerts reach you. Our travel health checklist covers medications and medical devices on the road.

Frequently Asked Questions

Is the Tandem Mobi insulin pump being recalled?

Yes. The FDA classified the Tandem Mobi recall as its most serious type on November 25, 2026. All Tandem Mobi pumps with Interoperable Technology are covered. Pumps on a specific serial number list should be replaced; others can stay in use with updated instructions.

How do I know if my Tandem Mobi needs to be replaced?

Compare your pump’s seven-digit serial number with the list in the FDA recall database entry (Z-0316-2027). The serial number is shown in the Tandem Mobi app under Settings, Pump, Pump Info. If you can’t confirm it, call Tandem at 1-877-801-6901.

Can I keep using my Tandem Mobi pump?

If your serial number is not on the replacement list, yes, as long as you respond to every occlusion alarm, keep app notifications on and have backup insulin ready. If it is on the list, request a replacement immediately and follow the same safety steps until it arrives.

How many people were hurt by the Tandem Mobi problem?

Tandem reported 1,248 serious injuries and no deaths linked to the false occlusion alarms as of October 28, 2026, according to the FDA.

Does the recall affect the Tandem t:slim X2 pump?

The FDA notice covers Tandem Mobi pumps with Interoperable Technology. The t:slim X2 is not named in this notice. Check with Tandem if you use a different model.

Sources: U.S. FDA, “Insulin Pump Recall: Tandem Diabetes Removes Tandem Mobi Insulin Pumps” (Nov. 25, 2026); FDA Medical Device Recall database, Z-0316-2027; CDC, Diabetic Ketoacidosis; MedTech Dive (Nov. 11, 2026); TechRepublic (Nov. 27, 2026). This article is for information only and is not medical advice. Talk with your diabetes care team before changing how you use your pump.

Pinny Surkis
Pinny Surkishttps://healthcarenewscenter.com/pinny-surkis/
Pinny Surkis is the founder and editor of Healthcare News Center, covering U.S. healthcare news for seniors, caregivers and people with disabilities. Based in Miami, he also runs Rank With Skill, a digital marketing agency.

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Share post:

Popular

More like this
Related

Every Reporting State Is Short on Home Care Workers, KFF Finds: Pay Rates and How Families Can Fill Gaps

All 49 states that answered KFF's 2026 Medicaid home care survey reported worker shortages, and 43 said they worsened. Here are state pay rates and the options families can use to fill care gaps.

FDA Approves Welgemda, First Treatment for Limb-Girdle Muscular Dystrophy: Trial Results and Who Qualifies

The FDA approved Welgemda (ribitol) on Nov. 23, 2026, the first treatment for any form of limb-girdle muscular dystrophy. Here is what the FORTIFY trial showed for walking and breathing, who qualifies, side effects and cost.

Nearly 1 in 3 Older Adults With Dementia Still Take High-Risk Sleep and Anxiety Drugs Linked to ER Visits

Nearly one in three community-dwelling older adults living with...

Thanksgiving Travel Health Checklist: How to Pack Medications and Plan for Mobility Needs

AAA expects 81.5 million Thanksgiving travelers. This guide helps older adults and caregivers pack medications, arrange assistance and plan for delays.